FDA Adverse Event Injury Summary report: N

IDENTITY XL DR

MDR report key: 4182214 · Received January 13, 2014

Report

Report Number
2017865-2014-03209
Event Type
Injury
Date Received
January 13, 2014
Date of Event
April 17, 2012
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DXY
PMA / PMN Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFO PROVIDED BY MFR, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PULSE GENERATOR EXHIBITED BACKUP VVI OPERATION. THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
25799 IDENTITY XL DR IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST. JUDE MEDICAL, INC., CRMD 5376

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention