FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 400 SYSTEM

MDR report key: 4181822 · Received August 4, 2014

Report

Report Number
3008642652-2014-02133
Event Type
Malfunction
Date Received
August 4, 2014
Date of Event
May 29, 2014
Report Date
July 8, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE OR WIRES EXPOSED) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT WAS UNABLE TO COMMUNICATE WITH A MONITOR. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A TRUNK CABLE PULLED FROM THE DISTRIBUTION NODE (DN). THE ROOT CAUSE FOR THE STRAINED CABLE WAS EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE PULLED CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(4) MALE PATIENT'S MOTHER CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A PULLED WIRE FROM THE PATIENT'S ELECTRODE BELT. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
453650 LIFEVEST WCD 400 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 47 YR