FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 400 SYSTEM
MDR report key: 4181822
·
Received August 4, 2014
Report
- Report Number
- 3008642652-2014-02133
- Event Type
- Malfunction
- Date Received
- August 4, 2014
- Date of Event
- May 29, 2014
- Report Date
- July 8, 2014
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
EVALUATION: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE OR WIRES EXPOSED) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT WAS UNABLE TO COMMUNICATE WITH A MONITOR. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A TRUNK CABLE PULLED FROM THE DISTRIBUTION NODE (DN). THE ROOT CAUSE FOR THE STRAINED CABLE WAS EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE PULLED CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(4) MALE PATIENT'S MOTHER CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A PULLED WIRE FROM THE PATIENT'S ELECTRODE BELT. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 453650 | LIFEVEST WCD 400 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |