PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-39594
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 18, 2014
- Report Date
- September 18, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST AND EXCESSIVE NO DELIVERY ALARM TEST. NO EXCESSIVE NO DELIVERY ALARM WAS NOTED. THE INSULIN PUMP WAS PROGRAMMED WITH MULTIPLE BOLUSES DELIVERIES AND ALL REGISTERED PROPERLY IN THE BOLUS HISTORY. NO BOLUS ANOMALY WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON THE DISPLAY WINDOW.
IT WAS REPORTED THE CUSTOMER WAS EXCESSIVELY RECEIVING NO DELIVERY ALARMS. HE STATED THAT HE HAD CHANGED OUT HIS INFUSION SET AND THE ALARM WAS RECURRING. THE CUSTOMER ALSO REPORTED THAT THE ALARM WAS NOT OCURRING DURING A MANUAL PRIME. HE ALSO REPORTED HIS INSULIN DID NOT LOOK CLEAR. THE CUSTOMER ALSO REPORTED RECEIVING A NO DELIVERY ALARM DURING A BOLUS. THE CUSTOMER COULD NOT RECEIVE HIS BOLUS AS A RESULT OF THE ALARM. TROUBLESHOOTING FOUND INSULIN WAS ABLE TO EXIT FROM THE TUBING DURING A FIXED PRIME, BUT THE INSULIN PUMP ALARMED NO DELIVERY. THE CUSTOMER WAS ADVISED THE ALARM MAY BE CAUSED BY AN INFUSION SET OR RESERVOIR OCCLUSION. CUSTOMER'S BLOOD GLUCOSE WAS 350 MG/DL. HE TREATED HIS BLOOD GLUCOSE WITH THE INSULIN PUMP. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 660027 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |