FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 4181497
·
Received October 17, 2014
Report
- Report Number
- 1416980-2014-36271
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 22, 2014
- Report Date
- September 22, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE FLOGARD INFUSION PUMP WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN AS PART OF PREVENTATIVE MAINTENANCE. THIS IS AN ANCILLARY SERVICE EVENT. A VISUAL INSPECTION AND A POWER ON SELF-TEST WERE PERFORMED. DURING THE VISUAL INSPECTION IT WAS IDENTIFIED THAT THE POWER CORD WAS DAMAGED. THE CAUSE OF THE DAMAGE IS UNKNOWN. THE POWER CORD WAS REPLACED; THE PUMP WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
Description of Event or Problem · 1
DURING ON-SITE SERVICE, THE BAXTER SERVICE TECHNICIAN FOUND THAT THE FLOGARD INFUSION PUMP HAD A DAMAGED POWER CORD. ADDITIONAL INFORMATION IS NOT AVAILABLE.