FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4181454 · Received October 17, 2014

Report

Report Number
2032227-2014-39555
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
September 18, 2014
Report Date
September 18, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING BOLUS. CUSTOMER'S BLOOD GLUCOSE READING WAS 189 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. IT WAS REPORTED THAT CUSTOMER WAS ABLE TO REWIND THE INSULIN PUMP. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660793 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 36 YR