PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4181316
·
Received October 17, 2014
Report
- Report Number
- 2032227-2014-39633
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- June 10, 2014
- Report Date
- June 10, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UNIT ALARMED A64 DURING SELF TEST DUE TO FAULTY Y1 AT RF BOARD. UNIT RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS. UNIT RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. (B)(6).
Description of Event or Problem · 1
IT WAS REPORTED THE CUSTOMER RECEIVED A FAILED SELF TEST ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 111 MG/DL. NO ADDITIONAL INFORMATION PROVIDED.