FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4181316 · Received October 17, 2014

Report

Report Number
2032227-2014-39633
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
June 10, 2014
Report Date
June 10, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNIT ALARMED A64 DURING SELF TEST DUE TO FAULTY Y1 AT RF BOARD. UNIT RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS. UNIT RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THE CUSTOMER RECEIVED A FAILED SELF TEST ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 111 MG/DL. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660370 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWL

Patients

Seq Age Sex Outcome Treatment
1