PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-39517
- Event Type
- Injury
- Date Received
- October 17, 2014
- Date of Event
- September 17, 2014
- Report Date
- September 18, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST AND EXCESSIVE NO DELIVERY ALARM TEST. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW.
THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4) SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE LEVELS. THE CUSTOMER STATED THAT HE WAS UNSURE WHAT WAS GOING ON WITH HIS INSULIN PUMP. THE BLOOD GLUCOSE READINGS WERE BETWEEN 500 TO 600 MG/DL SINCE THE PREVIOUS DAY, AND THE CUSTOMER NOTED HE HAD CHANGED HIS INSERTION SITE THE NIGHT BEFORE THE CALL. HE STATED THAT HIS BLOOD GLUCOSE LEVELS STAYED HIGH THROUGHOUT THE DAY, AND BY AFTERNOON IT WAS 289 MG/DL. AFTER EATING, THE BLOOD GLUCOSE READING CONTINUED TO LOWER TO ABOUT 220 MG/DL. HE ATE DINNER, PERFORMED A DUAL BOLUS, AND FOUND THE BLOOD GLUCOSE READING AT 179 MG/DL. BY MORNING THE NEXT DAY AROUND 4 AM, THE BLOOD GLUCOSE READING ROSE TO OVER 600 MG/DL, LOWERED TO 590 MG/DL, AND STAYED IN THE 500 MG/DL RANGE SINCE THEN. TROUBLESHOOTING FOUND THAT THE INSULIN PUMP FAILED THE HIGH PRESSURE TEST TWICE. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 660764 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |