FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 4181254 · Received August 4, 2014

Report

Report Number
3008642652-2014-02353
Event Type
Malfunction
Date Received
August 4, 2014
Date of Event
July 22, 2014
Report Date
July 30, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY CHARGER/MODEM WILL NOT POWER UP) WAS CONFIRMED. UPON EVALUATION, THE BATTERY CHARGER/MODEM WAS UNABLE TO POWER UP. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A DEFECTIVE POWER SUPPLY BRICK. THE ROOT CAUSE FOR THE DEFECTIVE POWER SUPPLY COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

A US DISTRIBUTOR CONTACTED ZOLL TO REPORT THAT BATTERY CHARGER/MODEM SN (B)(4) WAS UNABLE TO POWER UP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
453118 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA