FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 4180973 · Received October 17, 2014

Report

Report Number
1416980-2014-36241
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
September 23, 2014
Report Date
September 23, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN AS PART OF PREVENTATIVE MAINTENANCE. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION AND POWER ON SELF-TEST WERE PERFORMED. DURING VISUAL INSPECTION THE SAFETY CLAMP SHIM WAS FOUND TO BE DAMAGED; THE CAUSE OF THIS WAS UNKNOWN. TO CORRECT THE CONDITION, THE SAFETY CLAMP SHIM WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT SERVICE BY A FIELD SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED SAFETY CLAMP SHIM. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663005 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1