FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 4180960
·
Received October 10, 2014
Report
- Report Number
- 1720753-2014-08520
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- September 22, 2014
- Report Date
- October 10, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE COI NCELL BATTERY ON GPOS AND INTERCONNECT CABLE WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE SYSTEM WOULD NOT COMPLETELY BOOT UP. THERE IS NO REPORT OF DEATH OR SERIOUS INJURY ASSOCIATED WITH THE COMPLAINT. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 642844 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |