FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4180853
·
Received August 2, 2014
Report
- Report Number
- 3004209178-2014-89287
- Event Type
- Malfunction
- Date Received
- August 2, 2014
- Date of Event
- March 24, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MOTOR ERROR ALARM OCCURED DURING REWIND DUE TO A CORRODED MOTOR HOME SWITCH. DISPLACEMENT TEST COULD NOT BE PERFORMED DUE TO MOTOR ERROR ALARM. THE INSULIN PUMP HAD A SCRATCHED DISPLAY WINDOW, A CRACKED DISPLAY WINDOW CORNER, AND CRACKED RESERVOIR TUBE LIP. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM ON HIS OR HER INSULIN PUMP. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 452827 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |