FDA Adverse Event Malfunction Summary report: N

TRILOGY 202

MDR report key: 4180757 · Received October 15, 2014

Report

Report Number
2518422-2014-01797
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 23, 2014
Report Date
September 23, 2014
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE. DURING THE EVALUATION OF THE DEVICE AN ISSUE RELATED TO THE SENSOR BOARD WAS OBSERVED. THE SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
653774 TRILOGY 202 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1040000

Patients

Seq Age Sex Outcome Treatment
1