FDA Adverse Event Malfunction Summary report: N

AUTOLUBE-III

MDR report key: 4180728 · Received October 17, 2014

Report

Report Number
1045834-2014-14303
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
February 13, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
GEY
PMA / PMN Number
PK970530
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE DEVICE WAS DAMAGED AND NOT FUNCTIONING PROPERLY. THIS WAS ATTRIBUTED TO IMPROPER HANDLING OF THE DEVICE. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, TECHNICIAN FOUND SCREW MISSING, HOSE THIN AND RUBBERY, SEAL LEAKING, SWITCH WORN / DEFECTIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
662441 AUTOLUBE-III MOTOR, SURGICAL INSTRUMENT, AC-POWERED GEY DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1