FDA Adverse Event Malfunction Summary report: N

11 CM ANGLE ATTACHMENT QD11

MDR report key: 4180696 · Received October 17, 2014

Report

Report Number
1045834-2014-13732
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
August 21, 2012
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. (B)(4). DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, IT WAS NOTICED THAT THE DEVICE FAILED THE TEMPERATURE TESTING.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED FOR INSPECTION. DURING THE EVALUATION, THE TECHNICIAN REPORTED THE TOOL COULD NOT BE USED, AS THE EXTENSION CONNECTION WAS HEAVILY DAMAGED. THE GROUTING OF THE BEVEL WHEELS WAS INSUFFICIENT. THE BALL BEARINGS IN THE EXTENSION CONNECTION WERE DAMAGED. THIS WAS ATTRIBUTED TO NORMAL WEAR. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

.

Description of Event or Problem · 1

THE DEVICE RETURN FOR SERVICE DUE TO OVERHEATING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
662980 11 CM ANGLE ATTACHMENT QD11 MOTOR, DRILL, ELECTRIC HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1