FDA Adverse Event
Malfunction
Summary report: N
11 CM ANGLE ATTACHMENT QD11
MDR report key: 4180694
·
Received October 17, 2014
Report
- Report Number
- 1045834-2014-14415
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Report Date
- April 12, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE CONTROL UNIT DID NOT OPERATE. THIS WAS ATTRIBUTED TO NORMAL WEAR. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
COMPLAINT; ATTACHMENT HEATS UP. DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, TECHNICIAN FOUND THE FRONT BEARING BROKEN AND THE GUIDE TUBE ENLARGED. TECHNICIAN REPLACED FRONT BEARING AND GUIDE TUBE. THIS IS 1 OF 1 FOR COM-(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 662543 | 11 CM ANGLE ATTACHMENT QD11 | MOTOR, DRILL, ELECTRIC | HBC | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |