FDA Adverse Event Malfunction Summary report: N

11 CM ANGLE ATTACHMENT QD11

MDR report key: 4180694 · Received October 17, 2014

Report

Report Number
1045834-2014-14415
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
April 12, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE CONTROL UNIT DID NOT OPERATE. THIS WAS ATTRIBUTED TO NORMAL WEAR. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

COMPLAINT; ATTACHMENT HEATS UP. DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, TECHNICIAN FOUND THE FRONT BEARING BROKEN AND THE GUIDE TUBE ENLARGED. TECHNICIAN REPLACED FRONT BEARING AND GUIDE TUBE. THIS IS 1 OF 1 FOR COM-(B)(4).