ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-29578
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Report Date
- October 4, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WY, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/08/2014 WITH THE FOLLOWING FINDINGS: THE BATTERY COMPARTMENT WAS STRIPPED. THE BATTERY CAP WAS ALSO STRIPPED AND GOT STUCK ON THE BATTERY COMPARTMENT AND DID NOT SECURE TIGHTLY TO THE PUMP, SO A TEST CAP WAS USED FOR THE INVESTIGATION.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT THERE WAS A CASING/CONDITION ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.