FDA Adverse Event Malfunction Summary report: N

11CM ANGLE ATTACHMENT, BLACK MAX

MDR report key: 4180688 · Received October 17, 2014

Report

Report Number
1045834-2014-13311
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
June 12, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBE
PMA / PMN Number
PK831756
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE BEARINGS WERE DAMAGED AND THE DEVICE OVERHEATED. THIS WAS ATTRIBUTED TO NORMAL WEAR. MAINTENANCE WAS PERFORMED AND THE DEVICE WAS RETURNED TO THE CUSTOMER ON 06/27/2013.

Description of Event or Problem · 1

DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, TECHNICIAN FOUND BEARING WORN FROM USE, OVERHEAT. THIS IS REPORT 1 OF 1 FOR COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
662542 11CM ANGLE ATTACHMENT, BLACK MAX DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1