FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 4180648 · Received October 15, 2014

Report

Report Number
1720753-2014-08666
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
October 2, 2014
Report Date
October 15, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE HANDLES WERE EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE STEERING HANDLE AND CORRESPONDING WHEELS WERE DIFFICULT TO TURN. THERE IS NO REPORT OF DEATH OR SERIOUS INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
654734 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1