Summary report: N
TRILOGY 100
MDR report key: 4180626
·
Received October 15, 2014
Report
- Report Number
- 2518422-2014-01801
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- September 17, 2014
- Report Date
- September 17, 2014
- Manufacturer
- RESPIRONICS INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S LCD SCREEN AND SYSTEM BOARD TO LCD CABLE WERE REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFO ALLEGING A VENTILATOR'S LCD DISPLAY SCREEN WAS BLANK. THERE WAS NO HARM OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 655379 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC. | 1054260 |