FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 4180626 · Received October 15, 2014

Report

Report Number
2518422-2014-01801
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 17, 2014
Report Date
September 17, 2014
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S LCD SCREEN AND SYSTEM BOARD TO LCD CABLE WERE REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFO ALLEGING A VENTILATOR'S LCD DISPLAY SCREEN WAS BLANK. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
655379 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1054260

Patients

Seq Age Sex Outcome Treatment
1