FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 4180579
·
Received October 15, 2014
Report
- Report Number
- 1720753-2014-08630
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- September 25, 2014
- Report Date
- October 15, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AK, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE AND THE SOFTWARE WERE INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY WOULD NOT BOOT UP. THIS RESULTED IN AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY, WHICH COULD RESULT IN PROCEDURAL DELAYS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654575 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |