FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ 3.13 DUAL CHA
MDR report key: 4180552
·
Received October 15, 2014
Report
- Report Number
- 9615050-2014-05634
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- September 11, 2014
- Report Date
- September 17, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT DURING PREVENTATIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE HAD A CONTINUOUS AIR IN LINE ALARM ON CH B. NO ADDITIONAL INFO WAS PROVIDED. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, THE DEVICE DID NOT ALARM WHEN AIR WAS PRESENT IN THE TUBING ON CHANNEL B.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654563 | SYMBIQ 3.13 DUAL CHA | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |