FDA Adverse Event Malfunction Summary report: N

SYMBIQ 3.13 DUAL CHA

MDR report key: 4180552 · Received October 15, 2014

Report

Report Number
9615050-2014-05634
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 11, 2014
Report Date
September 17, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT DURING PREVENTATIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE HAD A CONTINUOUS AIR IN LINE ALARM ON CH B. NO ADDITIONAL INFO WAS PROVIDED. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, THE DEVICE DID NOT ALARM WHEN AIR WAS PRESENT IN THE TUBING ON CHANNEL B.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
654563 SYMBIQ 3.13 DUAL CHA 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA