FDA Adverse Event Malfunction Summary report: N

GEMSTAR PAIN MGE SNG

MDR report key: 4180539 · Received October 15, 2014

Report

Report Number
9615050-2014-05620
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 9, 2014
Report Date
September 10, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT THE DEVICE HAD A 09/006/003 (POWER ON MOTOR TEST) MALFUNCTION CODE AND A CRACK IN THE DISPLAY. THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, CORROSION DUE TO LEAKAGE FROM THE DISPOSABLE BATTERIES ON THE BATTERY DOOR, BATTERY COMPARTMENT, AND MIDDLE PRINTED CIRCUIT BOARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
655704 GEMSTAR PAIN MGE SNG 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA