FDA Adverse Event
Malfunction
Summary report: N
GEMSTAR PAIN MGE SNG
MDR report key: 4180539
·
Received October 15, 2014
Report
- Report Number
- 9615050-2014-05620
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- September 9, 2014
- Report Date
- September 10, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K060806
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT THE DEVICE HAD A 09/006/003 (POWER ON MOTOR TEST) MALFUNCTION CODE AND A CRACK IN THE DISPLAY. THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, CORROSION DUE TO LEAKAGE FROM THE DISPOSABLE BATTERIES ON THE BATTERY DOOR, BATTERY COMPARTMENT, AND MIDDLE PRINTED CIRCUIT BOARD.