FDA Adverse Event Malfunction Summary report: N

PINN STRAIGHT CUP IMPACTOR

MDR report key: 4180421 · Received October 17, 2014

Report

Report Number
1818910-2014-30157
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
October 13, 2014
Report Date
October 13, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LXH
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION OF THE RETURNED DEVICE CONFIRMS THE REPORTED EVENT OF END CAP HANDLE BREAKAGE. PREVIOUS EVALUATIONS BY DEPUY MATERIAL SCIENCE FOUND THE IMPACTORS WERE REPEATEDLY LOADED IN ROTATING BENDING FATIGUE BEYOND THE MATERIAL LIMIT RESULTING IN FATIGUE CRACK INITIATION AND PROPAGATION WITH MIXED-MODE OVERLOAD FINAL FRACTURE OF THE MATERIAL. NO EVIDENCE OF MANUFACTURING OR MATERIAL DEFECTS WAS OBSERVED THAT COULD CONTRIBUTE TO THE FAILURE OF THESE COMPONENTS. THE ROOT CAUSE IS ATTRIBUTED TO DEVICE OVERLOAD BASED ON SIMILAR EVALUATIONS. BASED ON THE PERFORMED INVESTIGATION, CORRECTIVE ACTION IS NOT NEEDED. CONTINUE TO MONITOR VIA SEP-419. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE END CAPS ARE BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
662818 PINN STRAIGHT CUP IMPACTOR HIP INSTRUMENT LXH DEPUY ORTHOPAEDICS, INC. 1818910 NB42638

Patients

Seq Age Sex Outcome Treatment
1