LOANER CORE SUMEX DRILL
MDR report key: 4180414
·
Received October 17, 2014
Report
- Report Number
- 0001811755-2014-03652
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 22, 2014
- Report Date
- September 22, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K112593
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING FAILURE ANALYSIS, THE REPORTED EVENT OF BIAS CURRENT MESSAGE WAS DUPLICATED AND CONFIRMED DURING SERVICE EVALUATION. THE PROBABLE CAUSE OF THE ISSUE WAS DETERMINED TO BE A DAMAGE OF THE CABLE ASSEMBLY.
Description of Event or Problem · 1
DURING A ROUTINE MAINTENANCE VISIT AT THE USER FACILITY, IT WAS REPORTED THAT THE LOANER CORE SUMEX DRILL DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. NO PATIENT INVOLVEMENT, NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.