FDA Adverse Event
Malfunction
Summary report: N
POSEY SITTER ELIT
MDR report key: 4180377
·
Received October 15, 2014
Report
- Report Number
- 2020362-2014-00333
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Report Date
- October 2, 2014
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION OF THE ALARM FOUND THERE IS A NOTE ON THE UNIT STATING "KEEP BEEPING". THE NOTE ON THE UNIT IS NOT CONFIRMED, THE UNIT DOES NOT SOUND CONTINUOUSLY WHEN IN USE WITH THE PULL MAGNET. THE PRE-RECORDED VOICE MESSAGE DOES NOT PLAY - ONLY A CLICKING SOUND IS HEARD WHEN WEIGHT IS REMOVED FROM THE SENSOR OR WHEN THE PULL MAGNET IS REMOVED WHEN IS DUE TO A DEFECTIVE VOICE CHIP. THE TONE PLAYS PROPERLY. NO PHYSICAL DAMAGE OBSERVED TO THE UNIT. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED THE ALARM IS NOT FUNCTIONING PROPERLY. CUSTOMER HAS LIMITED INFO. CUSTOMER COULD NOT PROVIDE THE DATE OF EVENT. NO PT INCIDENT OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 655373 | POSEY SITTER ELIT | KMI | J. T. POSEY CO. | 8345 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | MAGNET: 8345M |