FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4180348 · Received October 17, 2014

Report

Report Number
1416980-2014-36206
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
September 22, 2014
Report Date
September 22, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, A CLAMP BEING LEFT OPEN ON UNUSED LINES IS A KNOWN CAUSE OF THIS ALARM. USE ERROR AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE INSTRUCTS USERS TO MAKE SURE THAT ALL CLAMPS ON UNUSED FLUID LINES ARE CLOSED SECURELY AND TO CLOSE ALL CLAMPS WHILE PREPARING TO LOAD THE DISPOSABLE SET. FINALLY THE GUIDE WARNS THAT A SYSTEM ERROR 2240 CAN BE CAUSED BY UNCLAMPED, UNUSED SUPPLY LINES. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOMECHOICE (HC) DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET). THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING DWELL FOUR OF SIX OF PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, IT WAS DISCOVERED THAT AN UNUSED SUPPLY LINE CLAMP WAS LEFT OPEN. THE TECHNICAL SERVICES REPRESENTATIVE ASSISTED THE CARE GIVER IN CLEARING THE ALARM AND THE PATIENT WOULD CONTINUE THERAPY WITH ALL NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663292 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 57 YR HOMECHOICE