ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-07164
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 29, 2014
- Report Date
- October 1, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED. THE REMAINING BLADE PORTION WAS SCRATCHED, EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON GEN04 AN ERROR CODE 5 WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE STOP ACTIVATING AND DISPLAY AN ERROR CODE 5 IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN AN ERROR CODE 5 OR BLADE LOCKOUT LATER IN THE PROCEDURE. THE GENERATOR SYSTEM WILL REVERT TO STANDBY MODE, WHEN THE AUDIBLE ALARM (SOLID TONE) AND VISUAL ALARM INDICATOR APPEAR.
(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, SOLID TONE WITH ERROR CODE '5'. THE TIP WAS BROKEN DURING PRE-RUN TEST AFTER RE-INSTALLED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED. ONE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 663148 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |