FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 4180244 · Received October 15, 2014

Report

Report Number
1720753-2014-08590
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 24, 2014
Report Date
October 14, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SYSTEM SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.