FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 4180234 · Received October 15, 2014

Report

Report Number
1720753-2014-08591
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 24, 2014
Report Date
October 14, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE UPS WAS EVALUATED AN REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A UPS (UNINTERRUPTIBLE POWER SUPPLY) FAILURE ERROR MESSAGE INSTRUCTING USER TO RE-BOOT THE SYSTEM. SYSTEM FUNCTIONALITY WAS RECOVERABLE. THIS COULD CAUSE PROCEDURAL DELAYS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
653649 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900