FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 4180222 · Received October 15, 2014

Report

Report Number
1720753-2014-08607
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 24, 2014
Report Date
October 14, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE BOARDS AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM HAD A KV ON IN ERROR MESSAGE. THIS ERROR CAUSES THE SYSTEM TO SHUT DOWN. THIS RESULTING IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
653615 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment