FDA Adverse Event Malfunction Summary report: N

WATER PURIFICATION SYSTEM

MDR report key: 4180216 · Received October 17, 2014

Report

Report Number
3019131-2014-00028
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
September 19, 2014
Report Date
October 16, 2014
Manufacturer
MAR COR PURIFICATION
Product Code
FIP
PMA / PMN Number
K981085
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FACILITY REPORTED THEY EXPERIENCED A HIGH CONDUCTIVITY ALARM ON THEIR CWP RO SYSTEM (PERCENT REJECTION HAS DROPPED TO (B)(4)). UPON INSPECTION, THE CLINIC FOUND THE REVERSE OSMOSIS LOOP'S PRETREATMENT DEALKALIZER SCAVENGER HEAD WAS NOT FUNCTIONING CORRECTLY. THIS ALLOWED THE CWP RO TO INTAKE SOME BRINE CAUSING THE HIGH CONDUCTIVITY ALARM AND RO WATER PRODUCTION TO SHUT DOWN. THE SCAVENGER HEAD WAS BYPASSED AND THE WATER CONDUCTIVITY BEGAN TO FALL. 16 PATENTS WERE TREATMENT DELAYED. THERE WERE NO REPORTS OF PATIENT ILLNESS OR INJURY AS A RESULT OF THIS INCIDENT. THIS COMPLAINT WILL CONTINUE TO BE MAINTAINED WITHIN MAR COR PURIFICATION'S COMPLAINT SYSTEM.

Description of Event or Problem · 1

THE REVERSE OSMOSIS LOOP'S PRETREATMENT DEALKALIZER SCAVENGER HEAD WAS NOT FUNCTIONING CORRECTLY WHICH CAUSED THE CWP RO UNIT TO ALARM DUE TO HIGH CONDUCTIVITY. PATIENTS HAD TREATMENT DELAYED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
662623 WATER PURIFICATION SYSTEM REVERSE OSMOSIS SYSTEM FIP MAR COR PURIFICATION

Patients

Seq Age Sex Outcome Treatment
1