FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4180037
·
Received October 17, 2014
Report
- Report Number
- 1031452-2014-14661
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Report Date
- September 26, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER COMPLAINANT, UNIT ALARMING WITH LOW O2. INDICATION IS THAT SIEVE BED FILTERS (BOTTOM) ARE NOT SEATED PROPERLY ALLOWING SIEVE MATERIAL TO BECOME PULVERIZED AND BYPASS FILTER. SIEVE DUST IS MIGRATING INTO THE MANIFOLD VALVE AND MUFFLER. REFER TO (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 662654 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |