FDA Adverse Event Malfunction Summary report: N

HANA

MDR report key: 4180028 · Received October 2, 2014

Report

Report Number
4180028
Event Type
Malfunction
Date Received
October 2, 2014
Date of Event
September 29, 2014
Report Date
October 2, 2014
Manufacturer
MIZUHO ORTHOPEDIC SYSTEMS INC
Product Code
FWZ
Product Problem
Yes
Report Source
User Facility report
Reporter Location
NY, US

Narratives

Description of Event or Problem · 1

DURING PROCEDURE, LEFT FEMORAL LIFT FELL OFF OSI TABLE WHILE IN USE. BAR NOTED TO HAVE 4 SCREWS THAT WHERE SUPPOSE TO HOLD T EXTENSION OF LIFT TO TABLE, NONE OF WHICH WHERE STILL ATTACHED. NO OBVIOUS PT INJURY NOTED AT THIS TIME. STAFF WAS ABLE TO PUT BED BACK TOGETHER TO FINISH PROCEDURE. BED CHECKED BY BIOMED AFTER CASE COMPLETED.======================MANUFACTURER RESPONSE FOR HANA TABLE, (BRAND NOT PROVIDED) (PER SITE REPORTER).======================MANUFACTURER REPRESENTATIVE TO PROVIDE EDUCATION ON ROUTINE MAINTENANCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
614314 HANA TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED FWZ MIZUHO ORTHOPEDIC SYSTEMS INC * *

Patients

Seq Age Sex Outcome Treatment
1 68 YR NO OTHER THERAPIES