GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00950
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- October 14, 2014
- Report Date
- October 14, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER/GLASS CAP SHOWS A CIRCUMFERENTIAL FRACTURE AT THE FIBER/GLASS CAP FUSION ZONE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITS SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITS MILD CHAR. BASED ON THE DEVICE ANALYSIS, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.
IT WAS REPORTED THAT THE FIBER WAS FIRING STRAIGHT @ 66,167 JOULES (FORWARD FIRING). THE PROCEDURE WAS COMPLETED USING A SECOND FIBER. THE PATIENT OUTCOME IS LISTED AS ¿OK¿ REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 662783 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 413A |