FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 4180010 · Received October 17, 2014

Report

Report Number
2937094-2014-00947
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
September 17, 2014
Report Date
September 26, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER/GLASS CAP SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE DISTAL GLASS TIP OF THE FIBER WAS NOT RETURNED; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITS MODERATE CHAR; THE METAL CAP IS MELTED AT THE OUTPUT WINDOW LOCATION; THE GLASS CAP EXHIBITS SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW. BASED ON THE DEVICE ANALYSIS, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER 21 MINUTES WHILE USING THE SIDE-FIRING SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE OUTPUT BEAM WAS OBSERVED TO BE FORWARD-FIRING. THE PROCEDURE WAS COMPLETED USING A SECOND SURGICAL FIBER. PATIENT OUTCOME: "OK" - THERE WAS NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
662782 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 419A

Patients

Seq Age Sex Outcome Treatment
1