GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00947
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 17, 2014
- Report Date
- September 26, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER/GLASS CAP SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE DISTAL GLASS TIP OF THE FIBER WAS NOT RETURNED; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITS MODERATE CHAR; THE METAL CAP IS MELTED AT THE OUTPUT WINDOW LOCATION; THE GLASS CAP EXHIBITS SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW. BASED ON THE DEVICE ANALYSIS, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.
(B)(4).
IT WAS REPORTED THAT AFTER 21 MINUTES WHILE USING THE SIDE-FIRING SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE OUTPUT BEAM WAS OBSERVED TO BE FORWARD-FIRING. THE PROCEDURE WAS COMPLETED USING A SECOND SURGICAL FIBER. PATIENT OUTCOME: "OK" - THERE WAS NO INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 662782 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 419A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |