FDA Adverse Event Malfunction Summary report: N

IMPACT 73X PORTABLE VENTILATOR SYSTEM

MDR report key: 4178512 · Received October 16, 2014

Report

Report Number
2242630-2014-00007
Event Type
Malfunction
Date Received
October 16, 2014
Date of Event
August 22, 2013
Report Date
August 18, 2014
Manufacturer
IMPACT INSTRUMENTATION, INC.
Product Code
BTL
PMA / PMN Number
K051476
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

(B)(4). THIS SYSTEM WAS RETURNED FOR INTERNAL BATTERY PROBLEM. DURING THE INCOMING FUNCTIONAL TEST, THE ALARM FAILURE TRIGGER POINT WAS BELOW THE MINIMUM OF 10.6V. IT WAS DETERMINED THAT THE MAIN PCB WAS NOT FUNCTIONING CORRECTLY AND WAS REPLACED. THIS SYSTEM HAD PREVENTATIVE MAINTENANCE PERFORMED AND RETURNED TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
659187 IMPACT 73X PORTABLE VENTILATOR SYSTEM VENTILATOR BTL IMPACT INSTRUMENTATION, INC. 73X NA

Patients

Seq Age Sex Outcome Treatment
1 NONE REPORTED BY THE OPERATOR.