FDA Adverse Event Malfunction Summary report: N

IMPACT 73X PORTABLE VENTILATOR SYSTEM

MDR report key: 4178510 · Received October 16, 2014

Report

Report Number
2242630-2014-00004
Event Type
Malfunction
Date Received
October 16, 2014
Date of Event
July 3, 2013
Report Date
August 18, 2014
Manufacturer
IMPACT INSTRUMENTATION, INC.
Product Code
BTL
PMA / PMN Number
K051476
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

(B)(4). THIS SYSTEM WAS RETURNED TO IMPACT FOR PREVENTATIVE MAINTENANCE. DURING THE PREVENTATIVE MAINTENANCE SERVICE, IT WAS DETERMINED THAT THE OXYGEN REGULATOR WAS NOT PERFORMING TO SPECIFICATION, NOT HOLDING THE TIDAL VOLUME. THE O2 REGULATOR WAS REPLACED AND THE SYSTEM RETESTED AND FUNCTIONS NORMALLY. THIS SYSTEM WAS NOT SEEN BY IMPACT II FOR RECOMMENDED ANNUAL MAINTENANCE PRIOR TO THIS RETURN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
658331 IMPACT 73X PORTABLE VENTILATOR SYSTEM VENTILATOR BTL IMPACT INSTRUMENTATION, INC. 73X NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention NONE REPORTED BY THE OPERATOR