ENSEAL G2 ARTICULATING
Report
- Report Number
- 3005075853-2014-07118
- Event Type
- Malfunction
- Date Received
- October 16, 2014
- Date of Event
- September 29, 2014
- Report Date
- October 2, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K122797
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. UPON VISUAL INSPECTION UNDER MAGNIFICATION IT WAS NOTICED THAT THE UPPER JAW WAS SLIGHTLY DAMAGED AT THE RETENTION PIN INTERPHASE. RESULTING IN THE JAW BEING LOOSE AND DIFFICULTY IN OPENING AND CLOSING OF THE JAWS. HOWEVER THE DAMAGE WAS NOT SIGNIFICANT ENOUGH TO IMPEDE THE DEVICE FUNCTIONALITY. THE DEVICES WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED; AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). A PROBABLE CAUSE OF THE DAMAGE TO THE JAW COULD BE NOT ALLOWING THICK AND FIBROUS TISSUES TO DENATURE PRIOR TO I-BLADE ADVANCEMENT.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. DEVICE EVALUATION - NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THE JAW BROKE ON THE ENSEAL INSTRUMENT DURING AN UNKNOWN CASE. A SECOND INSTRUMENT WAS USED TO COMPLETE THE CASE WITH NO CONSEQUENCE TO THE PATIENT. .
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 657633 | ENSEAL G2 ARTICULATING | . | GEI | ETHICON ENDO-SURGERY, LLC. | NA | L4EX13 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |