FDA Adverse Event Malfunction Summary report: N

ENSEAL G2 ARTICULATING

MDR report key: 4177193 · Received October 16, 2014

Report

Report Number
3005075853-2014-07118
Event Type
Malfunction
Date Received
October 16, 2014
Date of Event
September 29, 2014
Report Date
October 2, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K122797
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. UPON VISUAL INSPECTION UNDER MAGNIFICATION IT WAS NOTICED THAT THE UPPER JAW WAS SLIGHTLY DAMAGED AT THE RETENTION PIN INTERPHASE. RESULTING IN THE JAW BEING LOOSE AND DIFFICULTY IN OPENING AND CLOSING OF THE JAWS. HOWEVER THE DAMAGE WAS NOT SIGNIFICANT ENOUGH TO IMPEDE THE DEVICE FUNCTIONALITY. THE DEVICES WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED; AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). A PROBABLE CAUSE OF THE DAMAGE TO THE JAW COULD BE NOT ALLOWING THICK AND FIBROUS TISSUES TO DENATURE PRIOR TO I-BLADE ADVANCEMENT.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. DEVICE EVALUATION - NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THE JAW BROKE ON THE ENSEAL INSTRUMENT DURING AN UNKNOWN CASE. A SECOND INSTRUMENT WAS USED TO COMPLETE THE CASE WITH NO CONSEQUENCE TO THE PATIENT. .

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
657633 ENSEAL G2 ARTICULATING . GEI ETHICON ENDO-SURGERY, LLC. NA L4EX13

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE