FDA Adverse Event Malfunction Summary report: N

NEWTON IQ W/STAY CYCLER SET

MDR report key: 4174544 · Received August 11, 2014

Report

Report Number
8030665-2014-00618
Event Type
Malfunction
Date Received
August 11, 2014
Date of Event
July 12, 2014
Report Date
July 12, 2014
Manufacturer
FRESENIUS MEDICAL CARE, REYNOSA MFG
Product Code
KDJ
PMA / PMN Number
K904806
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE PLANT'S INVESTIGATION.

Description of Event or Problem · 1

A PERITONEAL DIALYSIS (PD) PT REPORTED THAT HER CYCLER ALARMED FOR A MID-FILL PROBLEM DURING FILL 1. SHE WAS ADVISED TO RESET WITH NEW SUPPLIES. WHEN SHE WAS SETTING UP SHE FOUND A FLUID LEAK FROM THE SET DUE TO A TUBE BEING CUT BY A CRACK IN THE HOLDING TRAY. THE PT HAD ANOTHER TRAY AND SETUP WITH A NEW SET TO COMPLETE TREATMENT. THE SET WAS DISCARDED. DURING FOLLOW UP THE PT'S PD NURSE REPORTED THAT SHE HAS NOT HAD ANY SIGNS OF INFECTION. SHE WAS NOT TREATED WITH ANY ANTIBIOTIC. NO ADVERSE EVENT REPORTED AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
476261 NEWTON IQ W/STAY CYCLER SET KDJ FRESENIUS MEDICAL CARE, REYNOSA MFG 14BR08034

Patients

Seq Age Sex Outcome Treatment
1 67 YR NEWTON CYCLER