FDA Adverse Event Malfunction Summary report: N

UNIFY ASSURA DR CRT-D

MDR report key: 4173866 · Received October 15, 2014

Report

Report Number
2938836-2014-16689
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 9, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC WITH MULTIPLE EPISODES OF AMS. REVIEW OF EGM REVEALED FAR R-WAVE OVERSENSING ON THE ATRIAL CHANNEL. REPROGRAMMING WAS RECOMMENDED. THE PATIENT LEFT THE CLINIC WITH NO APPARENT R-WAVE OVERSENSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
654491 UNIFY ASSURA DR CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3357-40C NA

Patients

Seq Age Sex Outcome Treatment
1 71 YR