840 VENTILATOR
Report
- Report Number
- 8020893-2014-02354
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- September 15, 2014
- Report Date
- October 5, 2014
- Manufacturer
- COVIDIEN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB), BACKLIGHT INVERTER PRINTED CIRCUIT BOARD (PCBS) AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. (B)(4).
THE PRIOR MEDWATCH REPORT WAS SENT IN ERROR. THE EVENT INVOLVED IN THIS COMPLAINT IS NOT A REPORTABLE EVENT. PLEASE REPROCESS THIS REPORT NUMBER TO NOT BE A REPORTABLE EVENT.
COVIDIEN RECEIVED INFORMATION STATING THAT THE GRAPHIC USER INTERFACE (GUI) DISPLAY ON AN 840 VENTILATOR WAS UNREADABLE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654445 | 840 VENTILATOR | CONTINUOUS VENTILATOR | CBK | COVIDIEN | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |