ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-07100
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- July 7, 2014
- Report Date
- July 9, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH TISSUE PAD DETACHED BUT WITH EVIDENCE OF THE TISSUE PAD MATERIAL IN THE GROOVE SECTION OF THE CLAMP ARM. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND THE DEVICE DID ACTIVATE DURING FUNCTIONAL TESTING. BASED ON THE CONDITION OF THE TISSUE PAD, A PROBABLE CAUSE FOR THIS DAMAGE IS THAT THE CLAMP OF THE DEVICE MAY HAVE BEEN CLOSED AND THE INSTRUMENT ACTIVATED WITHOUT TISSUE PRESENT. CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. KEEP THE CLAMP ARM OPEN WHEN BACK-CUTTING OR WHILE THE BLADE IS ACTIVE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. THE RESULTING DAMAGE CONTRIBUTES TO THE REMOVAL OF THE PAD FROM THE CLAMP ARM. CLEANING OF THE PAD, NOT IN ACCORDANCE WITH THE IFU CAN ALSO RESULT IN REMOVAL OF THE PAD DURING USE.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TEFLON PAD OF THE ACE36E MELTED. CUSTOMER HAS CHANGED THE INSTRUMENT BECAUSE THERE WAS A RISK OF BURNINGS FOR THE PATIENT. THE CASE WAS COMPLETED BY USING ANOTHER ACE36E. CUSTOMER CANNOT PROVIDE ANY MORE DETAILS ABOUT THE CIRCUMSTANCES IN WHICH THE ISSUE OCCURRED. NO PATIENT CONSEQUENCE WAS REPORTED. ONE DEVICE IS BEING RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 656027 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |