OPTICROSS?
Report
- Report Number
- 2134265-2014-06339
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- September 18, 2014
- Report Date
- September 19, 2014
- Manufacturer
- BOSTON SCIENTIFIC - FREMONT (SUD)
- Product Code
- OBJ
- PMA / PMN Number
- K123621
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4).
DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. EVALUATION OF THE RETURNED DEVICE REVEALED THAT THE OUTER TELESCOPE TUBING WAS FOUND DETACHED FROM THE ANCHOR SEAL HOUSING. THE DEVICE CANNOT BE PROPERLY FLUSHED DUE TO THIS DETACHMENT. FLUID LEAKS FROM THIS JUNCTION. IMPEDANCE TESTING FINDS THE DEVICE IS CAPABLE OF IMAGING PROPERLY. FULL IMAGE CHARACTERIZATION TESTING CANNOT BE PERFORMED BASED ON THE DETACHED TELESCOPE TUBING. THIS EXPOSES THE IMAGING CORE AND COULD CAUSE IT TO WINDUP DURING TESTING. NO OTHER ISSUES OR DEFECTS WERE OBSERVED DURING PRODUCT ANALYSIS OF THE RETURNED DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE WAS UNABLE TO BE DETERMINED. (B)(4).
IT WAS REPORTED THAT AN OUTER SHEATH DETACHMENT OCCURRED. DURING PREPARATION FOR A PERCUTANEOUS CORONARY INTERVENTION (PCI), AN OPTICROSS¿ IMAGING CATHETER WAS SELECTED TO VISUALIZE AN UNSPECIFIED VESSEL. HOWEVER, IT WAS NOTED THAT THE OUTER SHEATH OF THE CATHETER BECAME SEPARATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT¿S CONDITION IS GOOD.
IT WAS REPORTED THAT AN OUTER SHEATH DETACHMENT OCCURRED. DURING PREPARATION FOR A PERCUTANEOUS CORONARY INTERVENTION (PCI), AN OPTICROSS¿ IMAGING CATHETER WAS SELECTED TO VISUALIZE AN UNSPECIFIED VESSEL. HOWEVER, IT WAS NOTED THAT THE OUTER SHEATH OF THE CATHETER BECAME SEPARATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT¿S CONDITION IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 655509 | OPTICROSS? | CATHETER, ULTRASOUND, INTRAVASCULAR | OBJ | BOSTON SCIENTIFIC - FREMONT (SUD) | H749518080 | 17151425 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |