FDA Adverse Event Injury Summary report: N

VERSAFITCUP CC TRIO ACETABULAR SHELL

MDR report key: 4173716 · Received August 19, 2014

Report

Report Number
3005180920-2014-00108
Event Type
Injury
Date Received
August 19, 2014
Manufacturer
MEDACTA INTERNATIONAL, SA
Product Code
LZO
PMA / PMN Number
K103352
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

DOCUMENT REVIEW: VERSAFITCUP CC TRIO SHELL CEMENTLESS SIZE 56: REF. 01.26.45.0056 / LOT 140081 ((B)(4) CUPS PRODUCED): ALL PARAMETERS WERE FOUND TO BE IN ACCORDANCE WITH THE SPECS VALID AT THE TIME OF MFG, INCLUDING WASHING AND STERILIZATION PROCEDURES. (B)(4) CUPS BELONGING TO THIS LIT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR INCIDENT. THE ISSUE WAS DISCUSSED WITH THE SURGEON AND EH ACKNOWLEDGED THAT INSTEAD OF REMOVING AND RE-REAMING BEFORE TRYING TO SEAT THE CUP, HE USED MORE FORCE COMBINED WITH A HEAVY MALLET. UNFORTUNATELY THE CUP DID NOT GAIN THE STABILITY HE HAD HOPED FOR AND IT WAS FINALLY REVISED. THE EVENT IS HIGHLY LIKELY NOT DEVICE RELATED.

Description of Event or Problem · 1

IMP REF #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498850 VERSAFITCUP CC TRIO ACETABULAR SHELL ACETABULAR SHELL 56 LZO MEDACTA INTERNATIONAL, SA

Patients

Seq Age Sex Outcome Treatment
1