FDA Adverse Event
Malfunction
Summary report: N
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
MDR report key: 4173713
·
Received October 15, 2014
Report
- Report Number
- 1061932-2014-02596
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- September 17, 2014
- Report Date
- September 17, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K081930
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ON (B)(4) 2014, A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE SYSTEM AND REPLACED VL218 RESOLVING THE ISSUE. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED CLEANER WAS NOT BEING PROPERLY REMOVED WHEN USING THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. THE ANALYZER GENERATED ERROR CONDITIONS OF VCS WASTE CHAMBER DRAIN INCOMPLETE, REQUIRED PARAMETER FOR DATA PLOT MISSING, AND VCSN LOW EVENT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER'S FACILITY TO EVALUATE THE SYSTEM. THERE WERE NO ERRONEOUS TEST RESULTS WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.