FDA Adverse Event Malfunction Summary report: N

UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM

MDR report key: 4173713 · Received October 15, 2014

Report

Report Number
1061932-2014-02596
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 17, 2014
Report Date
September 17, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K081930
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2014, A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE SYSTEM AND REPLACED VL218 RESOLVING THE ISSUE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED CLEANER WAS NOT BEING PROPERLY REMOVED WHEN USING THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. THE ANALYZER GENERATED ERROR CONDITIONS OF VCS WASTE CHAMBER DRAIN INCOMPLETE, REQUIRED PARAMETER FOR DATA PLOT MISSING, AND VCSN LOW EVENT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER'S FACILITY TO EVALUATE THE SYSTEM. THERE WERE NO ERRONEOUS TEST RESULTS WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
654125 UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA