EDWARDS SAPIEN XT TRANSCATHETER HEART VALVE, 23MM
Report
- Report Number
- 2015691-2014-02420
- Event Type
- Injury
- Date Received
- October 15, 2014
- Date of Event
- September 21, 2014
- Report Date
- September 29, 2014
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- NPT
- PMA / PMN Number
- P130009
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
PER THE INSTRUCTIONS FOR USE, CONDUCTION SYSTEM INJURIES (HEART BLOCK) WHICH MAY REQUIRE A PERMANENT PACEMAKER ARE POTENTIAL ADVERSE EVENTS ASSOCIATED WITH BALLOON AORTIC VALVULOPLASTY, DEPLOYMENT OF THE PROSTHETIC VALVE, AND THE OVERALL TAVR PROCEDURE. ACCORDING TO THE VALVE ACADEMIC RESEARCH CONSORTIUM (VARC) GUIDELINES, THE CLOSE ANATOMICAL RELATIONSHIP BETWEEN THE AORTIC VALVE COMPLEX AND THE BRANCHING ATRIOVENTRICULAR BUNDLE MAY PROVIDE AN EXPLANATION FOR THESE COMPLICATIONS OF THE TAVR PROCEDURE. ACCORDING TO LITERATURE REVIEW, AND AS DOCUMENTED IN AN EDWARDS TECHNICAL SUMMARY, ¿CONDUCTION DISTURBANCES/ HEART BLOCK¿, ATRIOVENTRICULAR CONDUCTION DISTURBANCES AFTER TAVR ARE ASSOCIATED WITH MANY PATIENT RELATED AND PROCEDURAL RELATED FACTORS, INCLUDING PRE-OPERATIVE CO-MORBID STATUS, THE DEGREE AND BULKINESS OF AORTIC VALVE AND ANNULAR CALCIFICATION, INTER-VENTRICULAR SEPTAL THICKNESS, PRE-EXISTING ELECTROCARDIOGRAM ABNORMALITIES, THE DEPTH OF PROSTHESIS IMPLANTATION, AND THE PROFILE OF THE IMPLANTED PROSTHESIS. UNLIKE CONVENTIONAL AVR, WHERE THERE MAY BE LOCALIZED TRAUMA DUE TO DECALCIFICATION OF THE ANNULUS AND/OR SUTURE PLACEMENT IN THE PROXIMITY OF THE AV NODE OR THE BUNDLES, TAVR MAY CAUSE CONDUCTION ABNORMALITIES THROUGH MECHANICAL IMPINGEMENT OF THE CONDUCTION SYSTEM BY THE PROSTHESIS. THE MECHANISMS OF THE DEVELOPMENT OF HEART BLOCK AFTER TAVR ARE WELL DOCUMENTED AND DESCRIBED IN THE LITERATURE. IT IS ALSO DOCUMENTED THAT PRE-EXISTING HEART BLOCK IS COMMON IN PATIENTS UNDERGOING TAVR OR SURGICAL AVR AND ANOTHER 4-6 % WILL DEVELOP POSTOPERATIVE HEART BLOCK, POTENTIALLY REQUIRING A PERMANENT PACEMAKER. IN THIS CASE, THE HEART BLOCK APPEARS TO BE RELATED TO THE MECHANISM DESCRIBED ABOVE. OTHER POTENTIAL CONTRIBUTING FACTORS ARE UNKNOWN AS NO RELEVANT MEDICAL HISTORY WAS PROVIDED. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED AT THIS TIME.
TWO DAYS POST PROCEDURE, PATIENT RECEIVED A PERMANENT PACEMAKER FOR COMPLETE HEART BLOCK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654124 | EDWARDS SAPIEN XT TRANSCATHETER HEART VALVE, 23MM | AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED | NPT | EDWARDS LIFESCIENCES | 9300TFX23A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention |