FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4173692 · Received October 15, 2014

Report

Report Number
2032227-2014-38341
Event Type
Injury
Date Received
October 15, 2014
Date of Event
September 15, 2014
Report Date
September 15, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

CUSTOMER REPORTED EXPERIENCING HIGH BLOOD GLUCOSE READINGS OF 435 MG/DL. TROUBLESHOOTING WAS PERFORMED AND THE DRIVE SUPPORT CAP AND ALL SETTINGS APPEARED TO BE NORMAL. LOW RESERVOIR ALARM WAS NOTED IN THE ALARM HISTORY. CUSTOMER STATED THAT THE RESERVOIR COMPARTMENT SMELLED LIKE INSULIN. SELF TEST WAS PERFORMED AND PASSED. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
654086 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAP

Patients

Seq Age Sex Outcome Treatment
1 38 YR