LOW PRESSURE BLACK MAX HANDPIECE
Report
- Report Number
- 1045834-2014-14359
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Report Date
- November 8, 2012
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBB
- PMA / PMN Number
- PK831756
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GR
- Reporter Occupation
- OTHER
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE OR UNCHANGED. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE BEARINGS COMPONENT WAS DAMAGED. THIS WAS ATTRIBUTED TO NORMAL WEAR. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
THE CUSTOMER REQUESTED THAT THE DEVICE BE RESTORED TO THE ORIGINAL INSTRUMENT. IT WAS REPORTED THAT THE HOSE HAS A HOLE, THE SECURITY LOCK IS DEFECTIVE, AND HAD LOW RPM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 653948 | LOW PRESSURE BLACK MAX HANDPIECE | MOTOR, DRILL, PNEUMATIC | HBB | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |