LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Report
- Report Number
- 3015876-2014-01217
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- September 4, 2014
- Report Date
- September 19, 2014
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- LDD
- PMA / PMN Number
- K063119
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). PHYSIO-CONTROL EVALUATED THE CUSTOMER¿S DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO-CONTROL REPLACED THE POWER SUPPLY ASSEMBLY. THEREAFTER PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING AND THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED POWER SUPPLY ASSEMBLY AND IT WAS OBSERVED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO PROCESS RESIDUE CONTAMINATION ON A FILTER, DESIGNATOR FL4, FROM THE POWER PCB ASSEMBLY. THE POWER PCB ASSEMBLY IS A COMPONENT LOCATED ON THE POWER SUPPLY ASSEMBLY.
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE REQUIRED SERVICING. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE DEVICE, PHYSIO-CONTROL OBSERVED THE DEVICE RESET ITSELF, OR LOCKED UP WHEN POWERED ON. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654901 | LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR | DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) | LDD | PHYSIO-CONTROL, INC | 20 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |