FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR

MDR report key: 4173638 · Received October 15, 2014

Report

Report Number
3015876-2014-01217
Event Type
Malfunction
Date Received
October 15, 2014
Date of Event
September 4, 2014
Report Date
September 19, 2014
Manufacturer
PHYSIO-CONTROL, INC
Product Code
LDD
PMA / PMN Number
K063119
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE CUSTOMER¿S DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO-CONTROL REPLACED THE POWER SUPPLY ASSEMBLY. THEREAFTER PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING AND THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED POWER SUPPLY ASSEMBLY AND IT WAS OBSERVED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO PROCESS RESIDUE CONTAMINATION ON A FILTER, DESIGNATOR FL4, FROM THE POWER PCB ASSEMBLY. THE POWER PCB ASSEMBLY IS A COMPONENT LOCATED ON THE POWER SUPPLY ASSEMBLY.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE REQUIRED SERVICING. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE DEVICE, PHYSIO-CONTROL OBSERVED THE DEVICE RESET ITSELF, OR LOCKED UP WHEN POWERED ON. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
654901 LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) LDD PHYSIO-CONTROL, INC 20

Patients

Seq Age Sex Outcome Treatment
1