FDA Adverse Event Malfunction Summary report: N

LOW PRESSURE BLACK MAX HANDPIECE

MDR report key: 4173594 · Received October 15, 2014

Report

Report Number
1045834-2014-14395
Event Type
Malfunction
Date Received
October 15, 2014
Report Date
October 15, 2012
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBB
PMA / PMN Number
PK831756
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE OR UNCHANGED. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE DEVICE WAS NOT FUNCTIONING PROPERLY. THIS WAS ATTRIBUTED TO NORMAL WEAR OF THE DEVICE. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

THE TECHNICIAN FOUND THAT THE DEVICE WAS TOO SLOW AND THAT THE AOV DID NOT STOP.THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
652984 LOW PRESSURE BLACK MAX HANDPIECE MOTOR, DRILL, PNEUMATIC HBB DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1